1.This device is an anesthesia agent analyzer used to measure EtCO₂, FiCO₂, RR, EtN2O, FiN2O, EtAA, FiAA.
2.This monitor is suitable for all kinds of animals and may be applied in general ward, including, but not limited to ICU, CCU or ambulance and so forth.
Main Unit’s Environment Requirement |
|
Working | Temperature: 5℃~50℃; Relative humindity: 0~95%; Atmosphere Pressure: 70.0KPa~106.0KPa |
Storage: | Temperature: 0℃~70℃; Relative humindity: 0~95%;Atmosphere Pressure: 22.0KPa~120.0KPa |
Power Specification |
|
Input Voltage: | 12V DC |
Input Current: | 2.0 A |
Physical Specification |
|
Main Unit | |
Weight: | 0.65Kg |
Dimension: | 192mm x 106mm x 44mm |
Hardware Specification |
|
TFT Screen | |
Type: | Colorful TFT LCD |
Dimension: | 5.0 inch |
Battery | |
Quantity: | 4 |
Model: | Rechargeable lithium battery |
Voltage: | 3.7 V |
Capacity | 2200mAh |
Working time: | 10 hours |
Recharging time: | 4 Hours |
LED | |
Patient Alarm Indicator: | Two colors: Yellow and Red |
Sound Indicator | |
Loudspeaker: | Play alarm voices |
Interfaces | |
Power: | 12VDC power socket x 1 |
USB: | MINI USB socket x 1 |
Measurement Specification |
|
Principle: | NDIR single beam optics |
Sampling Rate: | 90mL/min, ±10mL/min |
Initialization Time: | Waveform displaying in 20 seconds |
Range | |
CO₂: | 0~99 mmHg, 0~13 % |
N2O: | 0~100 VOL% |
ISO: | 0~6VOL% |
ENF: | 0~6VOL% |
SEV: | 0~8VOL% |
RR: | 2~150 bpm |
Resolution | |
CO₂: | 0~40 mmHg ±2 mmHg40 ~99 mmHg ±5% of reading |
N2O: | 0~100VOL% ±(2.0 vol% +5% of reading) |
ISO: | 0~6VOL%(0.3 vol% +2% of reading) |
ENF: | 0~6VOL%±(0.3 vol% +2% of reading) |
SEV: | 0~8VOL%±(0.3 vol% +2% of reading) |
RR: | 1 bpm |
Apnea Alarm Time: | 20~60s |
MAC value define |
|
|
|
Anaesthetic agents | |
Enflurane: | 1.68 |
Isoflurane: | 1.16 |
Sevflurane: | 1.71 |
Halothane: | 0.75 |
N2O: | 100% |
Notice | Desflurane’s MAC1.0 values differ with age |
Age: | 18-30 MAC1.0 7.25% |
Age: | 31-65 MAC1.0 6.0% |
*声明:深圳市美连医疗电子股份有限公司(简称“美连医疗”),以上内容中所有显示的注册商标、品名、型号等等,其所有权均为原持有者或原有厂商所有,本文仅用于说明美连医疗产品的兼容性,别无他意!以上所有信息仅供参考,切勿作为医疗机构或有关单位工作指南,否则,造成任何后果与本公司无关。